The touch-panel interface received a minor but significant update in 2021. The manual details these changes:

The 2021 manual explicitly states: “The CV-290 is not a sterilizer. It must be followed by either an AER (Automated Endoscope Reprocessor) or high-level disinfection per local protocol.” A flowchart on page 47 shows the exact sequence: olympus cv290 manual 2021

Bedside cleaning → Leak test → Manual washing → AER cycle → Drying storage → Pre-use check The touch-panel interface received a minor but significant


Purpose: This processor is designed for endoscopic imaging (GI, respiratory, and surgical procedures). It pairs with EVIS EXERA III series gastroscopes, colonoscopes, and duodenoscopes (e.g., GIF-HQ190, CF-HQ190). Bedside cleaning → Leak test → Manual washing

| Interval | Task | |----------|------| | Daily | Clean air/water and suction valves. Wipe external surfaces with 70% isopropyl alcohol. | | Weekly | Inspect water bottle O-ring (replace if cracked). Run automated channel flush cycle. | | Monthly | Calibrate color balance using Olympus test chart (part# MH-971). Check all foot pedal functions. | | Annually | Qualified biomedical engineer: Measure light intensity (should be >800 lux at scope tip). Check ground leakage current (<100 μA). Replace internal air filter. |